Imaging, Diagnosis, Prognosis |
Authors' Affiliations: 1 Department of Radiation and Cellular Oncology and 2 Center for Electron Paramagnetic Resonance Imaging In vivo Physiology, University of Chicago, Chicago, Illinois; and 3 Faculty of Biotechnology, Jagiellonian University, Krakow, Poland
Requests for reprints: Howard J. Halpern, University of Chicago Medical Center, MC1105, AMB ESB05a, 5841 South Maryland Avenue, Chicago, IL 60637. E-mail: howard{at}rover.bsd.uchicago.edu.
| Abstract |
|---|
|
|
|---|
1 mm and oxygen resolution of
3 Torr. The position of the Oxylite track was measured within a 2-mm accuracy using a custom stereotactic positioning device. A total of nine images that involve 17 tracks were obtained. Of these, most showed good correlation between the Oxylite measured pO2 and a track located in the tumor within the uncertainties of the Oxylite localizability. The correlation was good both in terms of spatial distribution pattern and pO2 magnitude. The strong correlation of the two modalities corroborates EPR imaging as a useful tool for the study of tumor oxygenation.
With the advent of modern radiation therapy delivery, intensity-modulated radiation therapy (11), it has become possible to subtly sculpt radiation dose distributions to avoid radiation sensitive anatomic structures. However, the region ascribed to the tumor-bearing tissue is treated homogeneously. This reflects our ignorance about regional variation in sensitivity within the tumor. If areas with high fractions of hypoxia could be defined, current technology could deliver substantially higher doses of radiation to those more hypoxic and resistant portions. (12) Present anatomic images of high resolution do not show changes in tissue physiology. A new, high-resolution, quantitative oxygen image is necessary to define the hypoxic regions to which these higher doses are delivered within the target volume. Such an oxygen image may provide at least an alternative and, because it is quantitative, a more repeatable and higher-resolution oxygen study than positron emission tomography scans to show response to therapy.
Because of its clinical importance, several techniques for measurements of oxygenation status in tissues and tumors of living animals and humans have been developed. Radiobiological hypoxic fractions (i.e., the proportion of clonogenic cells in a tumor, which are resistant to radiation due to hypoxia; ref. 13) provide a measure of tumor oxygenation. Selective retention of compounds that are reduced in hypoxic environments correlates with hypoxic fractions (14, 15) and can generate images of tumor hypoxia in living human subjects (16, 17). Oxygen concentrations in human tumors were measured with polarographic electrodes (8, 9). Point measurements of oxygen concentrations have been made with fluorescence quenching (18, 19) and particulate electron paramagnetic resonance (EPR) spectroscopy (20). Oxygen images have been obtained with fluorescence quenching (21), Overhauser enhanced magnetic resonance imaging (22, 23), 19F magnetic resonance imaging (24), and EPR imaging (25, 26). Noninvasive, three-dimensional tumor oxygen maps would improve clinical radiation treatment by enabling modulation of the radiation dose according to spatial information on tissue oxygenation.
Because each technique for hypoxia measurement emphasizes particular aspects of the cancer microenvironment, comparison of these techniques is crucial. Comparison both validates a new technique and helps understand more precisely what aspect of oxygenation physiology each is measuring. Hypoxic fractions have variably agreed (27) and disagreed (28) with oxygen electrode measurements. Disagreement may be due to the needle electrode insensitivity to necrosis. Comparisons of pooled measurements of oxygen electrodes and fiber optic fluorescence quenching have shown overall good cross validation (18, 19). Comparison of data from different imaging modalities or point measurements and images has been more difficult to obtain because of the difficulty in both spatial and temporal coregistration.
EPR oxygen imaging is a powerful new imaging technique that provides 3 Torr or better resolution in pO2 and better than 1-mm spatial resolution (26). The work presented here represents an effort to understand the relationship between pO2 images obtained using EPR spectroscopic imaging and Oxylite fiber optic point measurement of fluorescence quenching by oxygen. The difficulty in localizing the measurement in the EPR oxygen image was overcome by use of a stereotactic holder for the Oxylite fiber optic probe. Oxylite measurements were obtained in the EPR imager, just after an EPR tumor oxygen image with as little perturbation of the tumor as possible. The good agreement between Oxylite and the EPR oxygen image provides a strong cross-validation.
| Materials and Methods |
|---|
|
|
|---|
30 minutes) than that in other tissues.
Animal preparation. A total of nine tumors were used for the comparison of EPR imaging with Oxylite measurement. FSa cells (generously provided by Kathryn Mason, M.D. Anderson Hospital, Houston, TX) were injected i.m. in the right hind legs of 6-to 8-week-old C3H/HeN:Hsd (Harlan Sprague-Dawley, Indianapolis, IN) female mice so that the tumors grew to be
10 mm in diameter in 10 days. Animals were both immobilized and anesthetized for both Oxylite and EPR oxygen image measurements. This minimized animal movement during the procedures. Anesthesia involved i.p. ketamine (130 mg/kg) and xylazine (6 mg/kg) in 21.9 ± 0.8 g mice with upkeep of 20% this dose every half hour. In the EPR imager, the animal was immobilized by waterproof tape (WET-PRUF, Kendall, Mansfield, MA), and the tumor-bearing leg was placed in the 16-mm-diameter, 15-mm-thick loop of a loop-gap resonator. Tumor spatial stability and reproducibility were maintained by injecting hydrophilic vinyl polysiloxane dental impression material (GC Dental Products, Kasugai, Japan) under the leg to form a cast. The skin temperature, measured with a Physitemp (Clifton, NJ) digital needle probe thermometer, was kept at 33.5°C using opposed heating lamps. The spin probe rapidly accumulates in the mouse bladder (beginning within a minute after injection) and creates a source of confounding signal and artifact. To eliminate this, we flushed the bladder at 15 mL/h using a double-lumen urethral catheter and Harvard 22 syringe pump (Harvard Apparatus, Holliston, MA; ref. 29).
After placing the animal in the EPR imager and immobilizing the leg in the resonator, the EPR image was taken first. Immediately afterwards, without removing the animal, an Oxylite measurement was done. The interval between the beginning of imaging and the end of the Oxylite measurement was
2 hours. The complete experiment, including the animal preparation (
45 minutes), could be up to 3 hours long. We carefully maintained a steady level of anesthesia and monitored the breathing rate and temperature during the experiment to ensure that the oxygenation level of the tissues was similar for EPR imaging and Oxylite measurements.
All animal experiments were done according to the USPHS "Policy on Humane Care and Use of Laboratory Animals," and the protocols were approved by the University of Chicago Institutional Animal Care and Use Committee (ACUP No. 69681). The University of Chicago Animal Resources Center is an Association for Assessment and Accreditation of Laboratory Animal Careapproved animal care facility.
EPR imaging. Four-dimensional EPR images were obtained using a spectroscopic imager with an air-core magnet and Anderson type gradient coils (30) operating at 250 MHz and 9 mT as described previously (31). The Zeeman field was modulated at 4.98 KHz with overmodulation. We used an accurate line shape simulation that allows operation with high modulation amplitude (17 µT) to increase signal-to-noise ratio in images without sacrifice in the determination of the intrinsic line width (32, 33). The quality factor (Q) of the resonator was
280 without the tumor and 200 with the tumor installed. The power delivered to the resonator was 25 µW and well below the saturation level (200 µW).
EPR oxygen images were acquired and reconstructed using a four-dimensional tomographic spectroscopic imaging technique. This involves the acquisition of spectra with fixed gradients, referred to as projections. Each projection involved the acquisition of 256 field points scaled to uniformly populate the field interval determined by the gradient as has been described previously (34). The projections were subjected to a Gaussian filter whose width was four points and then subsampled to 64 points. A field of view of 3 cm was used. This determined our voxel size to be (3
2 cm/64) or 0.66 mm in the spatial direction. This voxel size determination has been discussed extensively elsewhere (26, 34).
Projections were obtained with 16 gradient amplitudes and 8 evenly spaced polar and azimuthal direction angles. An image required
45 minutes. The low signal-to-noise ratio at higher gradient projections were partially compensated for by signal averaging of multiple field sweeps, up to 10 averages for the highest gradient. Each sweep was obtained with 256 field points with acquisition time of 3 ms/point, 3 ms time constant, and 12-dB/octave filter using a SR830 lock-in amplifier (Stanford Research Systems, Sunnyvale, CA). The spatial field of view and the spectral field of view were 3 cm and 0.1 mT, respectively.
Spectroscopic image reconstruction used filtered back projection as described previously (26, 34, 35). All image voxels with spin probe amplitude
15% of the maximum spectral amplitude in the image were discarded. This 15% threshold was used to avoid fitting voxels with no signal or very poor signal. Small areas within several images had maximum intensities that did not survive this threshold. The oxygen values for these areas were set to zero. It should be noted that the Oxylite tracks were aimed toward the periphery of the tumor where such zero or very low signal regions were minimal.
The oxygen maps reconstructed and fitted from the images have a pO2 resolution of
3 Torr and a measured spatial resolution of
1 mm. Oxygen partial pressure resolution is limited by the accuracy of the spin probe spectral line width measurement. The actual source of the uncertainty is a complicated function of signal noise, sampling frequency, and the various forms of data filtering used in the reconstruction. In homogenous phantom imaging experiments, we have found that line width scatter is 0.16 µT, which is equivalent to 3 Torr via the oxygen versus line width calibration, which is discussed in ref. (26). Each image has a voxel size of 0.66 mm3. Thus, not only are the pixels shown in an image slice (0.66 mm)2, but also the slice is 0.66 mm thick.
Oxylite measurement. We used an Oxylite 2000 (Oxford Optronix, Oxford, United Kingdom) consisting of a 230-µm-diameter optical fiber for oxygen measurement and a thermocouple with fine wire leads. The oxygen probe tip (250-300 µm in diameter) has Ruthenium-III-(Tris)-chloride embedded in silicone polymer. It has a blue lightemitting diode that generates light pulses to induce fluorescence from the ruthenium luminophor. The oximetry is based on the principle that the lifetime of the fluorescent pulse is inversely proportional to the oxygen tension in the tip. A temperature sensor (T-type thermocouple with polyurethane coating) is attached in close proximity to the oxygen probe tip to correct for the temperature dependence of the fluorescence quenching. As delivered, the optical fiber can be moved into its Luer holder by as much as 4 mm. To fix the probe position more substantially to the holder, we applied the polysiloxane dental material between the probe and Luer fitting, which does not affect the calibration of the Oxylite (data not shown).4
Immediately after the EPR imaging while the animal was still in the EPR imager and under anesthesia, the sterotactic frame for location of the Oxylite probe tip was rapidly and easily installed. The skin of the tumor was punctured with a 21-gauge needle (0.8 mm diameter) mounted on the frame, which, in turn, guided the Oxylite probe tip. Once the skin was punctured, the fiber optic probe was stiff enough to proceed into the tissue in the direction defined by the stereotactic positioning device (Fig. 1
). The Oxylite probe was inserted into the puncture wound and then fully through the tumor to
1 cm depth. Occasionally, low pO2 readings were obtained at the surface opposite the introduction of the Oxylite due to trauma with the initial introduction of the probe (36). Several minutes were initially allowed to get a steady reading (19, 37). pO2 was measured after each 1-mm retraction of the probe, eventually to the initial insertion point (Fig. 1). Flexion in the probe gave an uncertainty in the lateral position of the Oxylite track to within 2 mm (see below). Variation in the location of the resonator in the direction of its axis (y direction in Fig. 1) in the gradient coordinate system gave a longitudinal uncertainty of 2 mm as well.
|
24 hours of use. At each experiment, we calibrated the probe by bubbling pure Argon and three gas mixes of N2 and O2 (3.0%, 5.6%, and 9.3%) through water at 37°C. The atmospheric pressure logged at an internet weather service (Weather Underground, http://wunderground.com), corrected for our altitude (183 m), was used for the calibration. Most of the tumors underwent two Oxylite track measurements that took up to 1 hour. We obtained a total of 17 track measurements from nine tumors.
Selecting the location of the Oxylite in the oxygen image. It was assumed that the Oxylite probe tip moved through the tumor parallel to the resonator axis, the y axis in Fig. 1A and subsequent figures. Although the position of the sterotactic frame can be controlled within 1-mm accuracy, the track of the probe tip in the tissue may deviate from the straight line parallel to the resonator axis. The advancing and retraction of the fiber optic probe can distort the longitudinal position of the tissue by compressing it or stretching it by
1 mm. It is difficult to localize the probe tip due to bending and curvature of the fiber optic probe. This curvature is shown in Fig. 2
, a high-resolution X-ray image of the mouse leg tumor immobilized as in the Oxylite-EPR image comparison. The overall uncertainty in the localization of the Oxylite track is
2 mm. This uncertainty was accounted for by searching the adjacent track space in the EPR oxygen image within this limit to find the track with the best match to the Oxylite pO2 profile. Given the high gradients in the oxygen distribution and the high spatial resolution of the EPR oxygen images, it was very difficult to locate the Oxylite probe tip to within the image resolution. We found that within the cylinder of uncertainty of the Oxylite probe tip location in the EPR image, we could reliably identify a sequence or "track" of EPR oxygen image values that agreed with the Oxylite values. Although this may seem circular, we felt that it was reasonable to use the track whose oxygen values best agreed with the Oxylite. We tested this by trying to find matches to an Oxylite track in distant portions of the tumor and failed to find such matches.
|
Additional data. Because of concern about the repeatability of these measurements, a second set of data was obtained. Ten tumors were imaged. In each tumor, two Oxylite tracks, for a total of 20 tracks, were obtained in a manner identical to that described above. These tracks were cast into the EPR oxygen image using the same methods described above, and the same analysis was done. The statistical characteristics of the two groups of measurements were compared by applying the Student's t test to the distributions of both Pearson product-moment correlation coefficients and mean absolute differences per track.
| Results |
|---|
|
|
|---|
The oxygen image of the anatomy inside the resonator, including the tumor, was obtained immediately before the Oxylite measurement and is shown in Fig. 3 . The oxygen image in Fig. 3 shows two slices parallel to the axis of the resonator. Their intersection is a line roughly parallel to the tibia. These are quantitative images and the color of the image corresponds directly to a voxel pO2 measured in Torr or mm Hg. The very bright, rich-colored regions in Fig. 3 represent highly oxygenated regions (bottom center in the top left slice) or superficial skin (about the surface of the image) with pO2 approaching 75 Torr. The darker blue-colored regions of the image show hypoxic tumor.
|
|
Figure 4A and C show two perpendicular planes in the oxygen image passing through the sequence of oxygen voxels shown with a black line. Examples of the pO2 readings both from Oxylite measurement and EPR image (the same slices shown in Fig. 3) are plotted in Fig. 4B. In this particular case, readings from adjacent tracks in the EPR oxygen image bracket the values from the Oxylite track. Referring back to Fig. 4A, one can see a region of high oxygen tension at the right end of the track. Moving to the left (toward negative Y location values), the track passes through very low pO2 values. It then passes just by the tip of a high oxygen protuberance, a region of substantial oxygen gradient. The high gradients seen in the oxygen image are reflected in the large variations in the Oxylite readings over 2 mm. The horizontal error bar placed at the bottom of Fig. 4B emphasizes the uncertainty in needle localization and the difficulty in extracting the point by point agreement between the two modalities.
Of the 17 tracks obtained from nine images, most showed good correlation between the Oxylite measured pO2 and the pO2 values from a track in the EPR oxygen image within the uncertainties of the Oxylite location. In most of the tracks where the product moment correlation R was not very high, one can still visualize the good spatial resemblance between the Oxylite and the oxygen image measurements. The Oxylite oxygen values with the oxygen values from closest EPR oxygen image track values are plotted in Fig. 5 . The corresponding Pearson product moment correlation coefficients are displayed in Table 1 . The variations in the Oxylite are reflected, for the most part, in the variations of the "tracks" from the EPR oxygen image. Moreover, as seen in Table 1 several of the tracks with low R values had very small average and median absolute differences. For example, the second to lowest mean absolute difference and the lowest median absolute difference (0.2 Torr) occurs in track 1 from September 21. This indicates very good agreement in values. However, the R value is only 0.28. A discrepancy in the opposite direction occurs between the two measures in the track from October 1 showing the largest mean and median absolute pO2 differences but an R value of 0.86.
|
|
|
Finally, in Fig. 6 , we display the distribution of the oxygenation values from both the Oxylite and the oxygen image measurements. The mean, mode, and median values for these distributions are tabulated in Table 1. As can be seen, the robust statistical variables (mode and median) values are virtually identical, whereas means, difficult to apply to such a skewed distribution, are close.
The correlation results and difference measures from the second set of data are also shown in Table 1. The mean of the Pearson product moment coefficients R, 0.64 from the first set and 0.67 from the second, are very similar and the difference is statistically insignificant (P = 0.71). The average of the mean absolute difference measures, 10.4 from the first set and 13.4 from the second set, are also not significantly different from one another (P = 0.2).
| Discussion |
|---|
|
|
|---|
Both EPR oxygen image slices in Figs. 3 and 4 and the data presented in Fig. 5 show the heterogeneity in the oxygenation of the mouse legs with implanted FSa tumors. The tracks passed through both tumor and muscle tissue. In 14 of 17 tracks shown in Fig. 5, extremely hypoxic areas of pO2 were found, often in the middle of the track. There are some well-oxygenated areas as well, with values of 40 Torr and sometimes more, most likely in the region of normal tissue.
Despite the overall good agreement, there are differences between the Oxylite tracks and the EPR oxygen image tracks. Note that seven of the images show small regions with large discrepancies between the Oxylite pO2 values and those from the EPR oxygen image. In all but one of these, the discrepancies occur at the most distal points in the Oxylite track. In five of the distal discrepancies, there are high EPR oxygen measurements, whereas the Oxylite measures low or zero oxygen values. There are at least two reasons why these differences might have occurred in this set of five tracks. The abscissa measures the distance from the entry point of the Oxylite into the tumor. (a) At the largest abscissa values, deepest in the tumor, the 200-µm glass fiber Oxylite probe is most likely to maximally deviate from the assumed straight line track. This could be as much as 2 mm as shown in Fig. 2, a very large distance in these oxygen images given the large oxygen gradients shown in Figs. 3 and 4. (b) Both of the September 30 images show very high oxygen concentrations at the surface of the tumor. These may be out of the range of reliability for the Oxylite.
Track 2 from September 17 shows a discrepancy between the Oxylite and the EPR oxygen image at the beginning of the track. Review of the EPR oxygen image shows very rapid variation of the oxygen values in tracks near the Oxylite. Slight misregistration can explain such discrepancies.
In track 1 from October 8, 2004, the trityl intensity was unusually low in the region of discrepancy. Thus, there is reason to suspect those data points in the EPR image in this very small region. This points out the richness of the information that is available from EPR images. There are other variables of the spectrum that measure the quality of the data. This allows assessment of situations where data are less than optimum or which subset of the data is not fully trustworthy.
| Conclusion |
|---|
|
|
|---|
| Acknowledgments |
|---|
| Footnotes |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. This article must therefore be hereby marked advertisement in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Note: M. Elas and K-H. Ahn contributed equally to this work.
4 Oxford Optronix, personal communication. ![]()
Received 2/28/05; revised 4/ 9/06; accepted 4/12/06.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
O. Levy, J. Relwani, T. Zaman, T. Even, B. Venkateswaran, and S. Copeland Measurement of blood flow in the rotator cuff using laser Doppler flowmetry J Bone Joint Surg Br, July 1, 2008; 90-B(7): 893 - 898. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Cancer Research | Clinical Cancer Research |
| Cancer Epidemiology Biomarkers & Prevention | Molecular Cancer Therapeutics |
| Molecular Cancer Research | Cancer Prevention Research |
| Cancer Prevention Journals Portal | Cancer Reviews Online |
| Annual Meeting Education Book | Meeting Abstracts Online |