| Abstract |
|---|
|
|
|---|
Experimental Design: To establish whether OCT could differentiate between the composite microstructural layers of the human airways and simultaneously determine in situ morphologic changes, using a bench-top OCT system, we obtained cross-sectional images of bronchi from 15 patients undergoing lung resections for cancer. All scanned sections underwent subsequent detailed histologic analysis, allowing direct comparisons to be made.
Results: OCT imaging enables characterization of the multilayered microstructural anatomy of the airways, with a maximum penetration depth up to 2 to 3 mm and 10-µm spatial resolution. The epithelium, subepithelial components, and cartilage are individually defined. The acquired OCT images closely match histologically defined patterns in terms of structural profiles. Furthermore, OCT identifies in situ morphologic changes associated with inflammatory infiltrates, squamous metaplasia, and tumor presence.
Conclusions: Our results confirm that OCT is a highly feasible optical tool for real-time near-histologic imaging of endobronchial pathology, with potential for lung cancer surveillance applications in diagnosis and treatment.
There have been tremendous advances in biomedical imaging technologies to assist lung cancer management. These include low-dose spiral computed tomography, positron emission tomography, and magnetic resonance imaging. Although these technologies are significantly better than conventional diagnostic approaches, they are inherently costly, potentially hazardous in terms of radiation dose (3, 4), and limited in availability nationwide. Furthermore, their sheer bulk prohibits use in immediate diagnostic/treatment environments, such as the bronchoscopy or surgical fields. In this respect, autofluorescence bronchoscopy (LIFE-bronchoscopy) enhances identification and diagnosis of in situ mucosal abnormalities, such as early cancerous changes (5); however, this method is still hampered by inadequate image resolution and tissue depth penetration (6). Other bronchoscopic technologies encompassing incorporation of high-frequency ultrasonography, such as EBUS, do achieve deep penetration of the airway tissue, but their spatial resolution is insufficient for clear demarcation of the microstructural profile and morphologic changes (79). In addition, due to small differences of acoustic impedance between the healthy and diseased tissues, the imaging contrast offered by EBUS is inherently low.
Optical coherence tomography (OCT) is a rapidly evolving imaging modality, which is noninvasive and noncontact; unlike ultrasonography, OCT does not require a transducing medium (10). Whereas analogous to B-mode ultrasonography, rather than using a sound signal, OCT delivers near IR light waves to the imaged site through a single optical fiber. This light reflects off the internal microstructural layers within the scanned tissue, allowing micron-scale resolution pick-up of the normal anatomy and any in situ morphologic aberrations. Signal processing involves low coherent interferometry, which is basically the analysis for reflected light waves from the internal tissue microstructures (10, 11). The coherence length of the light source determines the longitudinal resolution. With the appropriate light source, OCT imaging can reliably produce a 5- to 15 µm resolution compared with 150 µm for high-frequency ultrasonography (9). Thus, availability of high-tech OCT beam delivery instruments would constitute a major advance in lung cancer surveillance; importantly, its compactness makes OCT suitable for use in bronchoscopy and during thoracic surgery.
We have already shown OCT can provide near-histologic images of the different microstructural profiles along the tracheobronchial tree and detection of endobronchial foreign body deposition (12, 13). In this present study, we test OCT imaging in freshly excised lungs of patients undergoing resection surgery for lung cancer; we explore OCT's capability to identify in situ inflammatory and neoplastic aberrations within the bronchial airway and determine whether OCT images reflect conventional histologic analysis.
| Materials and Methods |
|---|
|
|
|---|
2 mW; Fig. 1). The light source used yields a
10-µm axial resolution in lung tissue (12, 13), with the mean refractive index of bulk airway mass taken as 1.4. The output from the light source is coupled into a 2 x 1 single mode optical fiber, then further split by a 50:50 optical fiber coupler. Fifty percent of the light is directed to the reference arm of the interferometer, where a rapid double-pass scanning system is employed to modulate the interference signal and provides the optical path length scanning (14). The residual light (
1 mW) is directed towards the resected lung sample, with a focusing optics. A high-resolution motorized translation stage accurately controls mirror movement. Light backscattered from the imaged sample is combined with light reflected from the reference arm. Polarization controllers are used in both arms to achieve the maximum obtainable interference fringe visibility. The system employs a balanced detector scheme to minimize the fluctuation noise arising from the light source. The transverse resolution measures 16 µm, limited by the numerical aperture of the lens delivering light onto the sample, and the optical frequency of the incident light as in conventional microscopy. The system signal-to-noise ratio measures
100 dB, using a 4 OD neutral density filter. The scanning rate of the current system is 200 A scans per second.
|
Histologic analysis. The scanned tissue sections were then fixed in 10% buffered formalin for 48 hours and subjected to standard paraffin embedding processing. Sections
5µm thick were cut at the marked tissue sites and stained with standard H&E. Tissue slide examination and micrographs were done with a Nikon Eclipse 80i (Nikon, Melville, NY) and recorded with a Nikon digital Net Camera DN100 on a Nikon Eclipse E600 light microscope (Nikon), respectively. OCT images were directly compared with histologic sections.
Image and data reproducibility. To avoid observer bias, analytic comparison of the recorded tomograms and measurement of structural dimensions were carried out independently by two histopathologists (blinded to one another) and the interobserver variability assessed. Each observer separately matched anatomic profiles on OCT tomograms with corresponding histology images. In addition, thickness measurements were done of the epithelium, cartilage plates, and the distance between these two layers in three separate areas of each OCT tomogram and corresponding histologic section; a total of 10 representative images for each patient were analyzed. P < 0.05 was accepted as statistically significant.
| Results |
|---|
|
|
|---|
OCT characterization of disease-free bronchial airway. We used corresponding histology sections of the scanned macroscopically disease-free airways for validation of OCT images. Figure 2A shows the anatomy of a human healthy airway with its characteristic multilayered profile. OCT images precisely delineate in cross-section the following anatomic components: epithelium (E), lamina propria (LP), smooth muscle (SM), mucus glands (G) and cartilage (C; Fig. 2B). The transition among these microstructural layers is well defined and closely mirrors the layered appearance profiled on the histology image. The demarcation of the epithelium, mucous gland ducts, and cartilage is specifically mapped by OCT; the lamina propria and submucosal structures, while easily recognizable, at times seem to be less well demarcated. This variation in OCT definition across the different layers of the airway wall could be explained by the presence of a higher nuclear density within structures, such as the epithelium and cartilage, reflected in enhanced reflectivity signals compared with adjacent surrounding tissues. Thus, the relatively higher refractive index of a particular structure results in sharper OCT image interpretation. For example, the comparatively denser extracellular matrix of cartilage decreases scattering of incident light and thus reflects as a dark region on the OCT tomogram. The connective tissue layer, including smooth muscle beneath the epithelium, is also visualized, but we could not differentiate between smooth muscle bundles and blood vessels.
|
Identification of in situ inflammatory and neoplastic pathology. The presence of chronic intense inflammation tends to homogenize tissue and disrupts tissue boundaries histologically (Fig. 3A). Consequently, inflamed tissue results in a less well defined OCT image, where only the epithelium and lamina propria are clearly delineated (Fig. 3B). However, whereas the epithelium is seen to remain intact, the relatively deeper border of lamina propria merges indistinguishably into the submucosa with increasing infiltration from inflammatory cells. Under such circumstances, as relative differences in refractive index between structural components decrease, it becomes increasingly difficult to distinguish cartilage on these OCT images (Fig. 3). In addition, inflamed tissue is more vascular than normal, thereby further reducing the imaging capability due to strong absorption of light at 1,300 nm. However, even at a resolution of 10 µm, OCT imaging captures the remodeling present within inflamed airways as visualized on histologic analysis.
|
|
|
The relative dimensions of structural components on OCT images is comparable with histologic analysis. The measured thickness on OCT images of the epithelium and cartilage is 100 ± 25 and 450 ± 15 µm, respectively, whereas the intervening distance from epithelium to cartilage is 250 ± 28 µm compared with their histologic quantification of 84 ± 21, 378 ± 30, and 210 ± 42 µm, respectively. Quantitative assessment was done on three separate areas of each corresponding OCT tomogram and histology section from macroscopically healthy airways; a total of 10 representative images for each patient were analyzed. There is no significant difference between patients in terms of thickness of healthy bronchial mucosa components (P > 0.5). A consistent shrinkage of
16% was observed in microstructural component thickness on histologic images compared with OCT. This is due to inevitable post-fixation tissue desiccation. In addition, assigning an average refractive index of 1.4 across the whole airway profile may contribute to a slight overestimation/underestimation of measurements as done on OCT images (e.g., cartilage has an inherent refractive index of 1.51).
| Discussion |
|---|
|
|
|---|
OCT is analogous to optical "ultrasound," delivering near IR light waves (rather than sound) to the imaged tissue through a single optical fiber. As such, OCT represents a technological shift from another current methods employed in lung cancer screening. By using variations in optical backscatter within the imaged tissue, OCT can effectively differentiate normal airway profiles from regions bearing in situ pathologic aberrations. As the ordered multilayered anatomy of the healthy airway is disrupted by structural remodeling consequent on development of inflammation and/or neoplasia, OCT tomograms reflect bland images, which ignore the normal microstructural boundaries. Inflammation is characterized by a loss of depth beyond the epithelial and lamina propria layers on the OCT image. This partly results from the presence of infiltrating inflammatory cells, such as lymphocytes, which have a high nuclear density, thereby modulating the refractive index of the imaged bulk tissue. In addition, enhanced vascular supply within the inflamed region contributes to reduced imaging capability, as whole blood induces strong attenuation of light. This originates from the absorption of light by hemoglobin and the scattering properties of RBC (15). In contrast, invasive neoplastic changes seem as featureless images on OCT, reflecting total loss of local architecture characteristic of tumor infiltration seen on subsequent histologic examination. Similar images have been reported in the gut tumors (16, 17). Continuing developments in OCT engineering will include measures to overcome these pitfalls in future as differential tissue contrast information is gleaned from imaging various states of human airway pathology.
We acknowledge our current OCT system is limited in its ability to define individual cellular profiles, and that this would be a significant step forward to enable OCT imaging to revolutionize lung cancer surveillance. Developments will need to include faster data acquisition and processing for in vivo real-time recording and enhancement of resolution by use of broad-band light sources, OCT-multiphoton microscopy, or polarization-sensitive OCT (18). Inevitably, such developments will need to be finely balanced against constraints placed by the particular in vivo milieu in which OCT is being used and the complex tissue profile. Thus, whereas improvements in resolution <2 µm have already been achieved (19), it may be difficult to enhance imaging depth beyond 2 to 3 mm within the same OCT system. We are encouraged by the ability of our current OCT system to identify areas of squamous metaplasia along an airway; specifically as dysplastic changes, which precede invasive neoplasia often by years, show a progressive increase in the thickness of epithelium, that usually parallels the grade of dysplasia. Aside from neoplastic changes, the bronchial airway epithelium can also undergo remodeling in thickness and morphology associated with development of chronic airways diseases (20). Thus, the availability of such "optical pathologic fingerprints" through improvements in OCT technology would enable diagnosis of the early distinct cellular hyperplastic/metaplastic changes associated with lung cancer and chronic airway diseases.
Although the clinical use of OCT imaging has been tested in a variety of biological tissues (2124), its use in lung screening is still in its early stages. We foresee OCT imaging proving invaluable within two scenarios in thoracic medicine. The urgent need for better visualization of the endobronchial tree to monitor in situ mucosal aberrations and early diagnosis of lung cancer necessitate the exploitation of an OCT catheter design for incorporation into conventional bronchoscopy. Lessons may be obtained from other medical fields; recent reports of a small OCT catheter consisting of a rotary optical fiber joint for intravascular imaging (25) and the MEMS-based OCT endoscope for imaging urinary tracts (26) could be exploited in future designs of an OCT bronchoscopic system for routine in vivo screening. Obviously, engineering of such a flexible system would need to overcome the intricate changing anatomic profile of the human tracheobronchial tree and other issues, such as blood flow changes. However, an immediate potential for in vivo surface and near-surface imaging application for our OCT system is within thoracic surgery. As complete lung tumor removal offers best long-term survival for cancer patients, there is an urgent need for safe noninvasive tools for use at operation, to enable real-time near-histologic information on early cancerous changes within the lung/thoracic cavity and to determine precisely resection margins. Currently, tissue disease status at operation is done by gross examination and, in some cases, cytologic/histologic assessments involving significant laboratory time, expense, and wait delay. Optimizing the current OCT system for feasible use during lung cancer surgery would enhance patient operability, preserve tumor-free lung in the compromised patient, and guide implementation of any adjunct interventions (27, 28).
| Acknowledgments |
|---|
| Footnotes |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. This article must therefore be hereby marked advertisement in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Received 2/ 3/05; revised 8/22/05; accepted 11/ 7/05.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
H. O. Coxson, B. Quiney, D. D. Sin, L. Xing, A. M. McWilliams, J. R. Mayo, and S. Lam Airway Wall Thickness Assessed Using Computed Tomography and Optical Coherence Tomography Am. J. Respir. Crit. Care Med., June 1, 2008; 177(11): 1201 - 1206. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Lam, B. Standish, C. Baldwin, A. McWilliams, J. leRiche, A. Gazdar, A. I. Vitkin, V. Yang, N. Ikeda, and C. MacAulay In vivo Optical Coherence Tomography Imaging of Preinvasive Bronchial Lesions Clin. Cancer Res., April 1, 2008; 14(7): 2006 - 2011. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Thiberville, S. Moreno-Swirc, T. Vercauteren, E. Peltier, C. Cave, and G. Bourg Heckly In Vivo Imaging of the Bronchial Wall Microstructure Using Fibered Confocal Fluorescence Microscopy Am. J. Respir. Crit. Care Med., January 1, 2007; 175(1): 22 - 31. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Cancer Research | Clinical Cancer Research |
| Cancer Epidemiology Biomarkers & Prevention | Molecular Cancer Therapeutics |
| Molecular Cancer Research | Cancer Prevention Research |
| Cancer Prevention Journals Portal | Cancer Reviews Online |
| Annual Meeting Education Book | Meeting Abstracts Online |